1. Introduction of Beating and Freeness Tester
Beating and Freeness Tester is used for the determination of the drainability of a pulp suspension in water in terms of the Schopper-Riegler (SR) number.
The Schopper-Riegler test is designed to provide a measure of the rate at which a dilute suspension of pulp may be dewatered. It has been shown that the drainability is related to the surface conditions and swelling of the fibres, and constitutes a useful index of the amount of mechanical treatment to which the pulp has been subjected
In principle, this method is applicable to all kinds of pulp in aqueous suspension. However, in practice, the Schopper-Riegler test provides acceptable results only if a sufficiently dense mat of fibres is formed on the wire screen. For this reason, the test is not recommended for some extremely short-fibred pulps, such as those from well-beaten hardwoods, as most of the fibres will pass through the wire screen, resulting in anomalous reduction of the SR number. The most reliable results are obtained within the range of 10 to 90 SR number.

2. Advantages
non-corrosive
low maintenance
Easy to use and clean
Stainless steel
Robust design
3. Principle
Draining, through a fibre mat formed during the test on a wire screen, of a given volume of pulp in aqueous suspension into a funnel provided with a bottom and a side orifice. Collection of the discharge from the side orifice in a measuring cylinder, graduated in Schopper-Riegler numbers.

4. Specifications
| DJD-01 | |
| Measuring Range | 1~100 SR |
| Division Value of Measuring Cylinder | 1 SR |
| Orifice Draining Time | 149±1s |
| Residual Volume | (7.5~8.0)ml |
| Sealing Cone Raising Speed | 100±10mm/s |
| Dimension | 450*250*1200mm |
| DJD-02 | |
| Measuring Range | 1~100 SR |
| Operation | Auto Release |
| Sealing Adjustment | Auto Sealing |
| Side Orifice Adjustment | Adjustment button equipped, random setting of residual volume |
| Orifice Draining Time | 149±1s |
| Residual Volume | (7.5~8.0)ml |
| Sealing Cone Raising Speed | 100±10mm/s |
| Dimension | 410*360*760mm |
| DJD-03 | |
| Measuring Range | 1~100 SR |
| Display | Digital, drainage curve analysis |
| Resolution | 0.1 SR |
| Operation | Auto Release |
| Sealing Adjustment | Auto Sealing |
| Side Orifice Adjustment | Adjustment button equipped, random setting of residual volume |
| Orifice Draining Time | 149±1s |
| Residual Volume | (7.5~8.0)ml |
| Sealing Cone Raising Speed | 100±10mm/s |
| Dimension | 410*360*760mm |
5. Standards
ISO 5267-1, GB/T 3332, QB/T 1054

6. FAQs
What is the Schopper-Riegler number scale?
The scale on which a discharge of 1 000 ml corresponds to a SR number of zero and zero discharge to a SR number of100.
What is ‘stock’ in pulp beating freeness test?
aqueous suspension of disintegrated pulp.
If you have purchase needs, please contact us and Cell Instruments will give you satisfactory products.
この製品に関する詳細情報をご希望の場合は、お気軽にお問い合わせください。その他の人気商品をご紹介します: Tissue Thickness Tester